It was reported 'the functionality of the endoscopic gatroc release system (egr) blade was tested before the device was inserted into the guide provided by the kit.The device worked properly when tested.Once the device was inserted into the guide, during surgery, there was some movement as the surgeon tried to align the blade with the sit in the guide.Once it was properly aligned, the surgeon attempted to deploy the blade which worked but the surgeon was unable to recess the blade back into the device when needed.The blade was stuck in the deployed position.There was no patient injury or adverse consequence alleged.The surgery was delayed for 5 minutes due to this issue'.
|