Result: the catheter is ovalized approximately 1.0, 2.0, 17.0, 39.0, 72.0, and kinked 90.0 cm from the distal tip.Conclusion: the complaint has been evaluated.The complaint indicates that suction during the aspiration was not smooth.The catheter was removed and several kinks were found.Evaluation of the returned device confirmed several ovalizations and a kink along the length of the distal catheter shaft.This damage is commonly observed when the product is mishandled during preparation for use or during manipulation in the patient.These devices are 100% inspected for damage during processing.This damage was not noted upon inspection of the device out of the packaging before insertion into the patient.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|