Model Number 9900
Device Problems Device Operates Differently Than Expected (2913); Inappropriate or Unexpected Reset (2959)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/12/2014
Event Type  malfunction  
Event Description
The fse reported battery and image processor failures which resulted in intermittent no boots, reboots and communication failures.This resulted in an intermittent, recoverable loss of imaging functionality.There is no report of injury or death.
 
Manufacturer Narrative
A ge service representative performed an onsite investigation.The cmos battery and image processor board were evaluated and replaced.The system software and configuration files were reloaded.The voltage on ffb board was readjusted.The output pins on generator ps1 was cleaned.The system was tested and found to be working as intended and returned to service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
9900
Type of Device
FLUOROSCOPIC X-RAY
Manufacturer (Section D)
GE OEC MEDICAL SYSTEMS (SLC)
384 wright brothers dr.
salt lake city UT 84116
Manufacturer (Section G)
GE OEC MEDICAL SYSTEMS (SLC)
384 wright brothers dr.
salt lake city UT 84116
Manufacturer Contact
384 wright brothers dr.
salt lake city, UT 84116
8015364952
MDR Report Key3979093
MDR Text Key4753974
Report Number1720753-2014-04620
Device Sequence Number1
Product Code JAA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/02/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9900
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/12/2014
Initial Date FDA Received06/02/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-