The patient reportedly did not fuse and the patient developed stenosis at the adjacent level.The original surgery l3-5 was performed on (b)(6) 2012.The fracture occurred some time ago but, the patient did well prior to developing stenosis at l1-2 &l3-4.The fractured 5.5x40 screw was removed, replaced with a 6.5x45 screw and the fusion was extended to include l1 & l2.The broken screw was retained by the hospital and will not be returning however, the manufacturing records were reviewed and no discrepancies were found.This report is being filed as a precaution as the incident is not believed to be related to the product.The root cause of the fracture is likely the pseudoarthrosis.These screws are intended to serve as an adjunct to fusion and in the absence of fusion 2 years post-operatively, a fracture of any of the screws is possible.Part retained by hospital.
|