A laboratory tech at (b)(6) hosp located at (b)(6) reported cutting their finger on a vial broken during handling.The laboratory tech was wearing gloves at the time of the incident, no sutures were required.The cause of the vial breakage is undetermined.The complaint also reported that the laboratory tested the remaining vial contents using vitros 5600 and obtained results (b)(6) for (b)(6).The customer generated (b)(6) results when performing secondary testing using the (b)(6) test.Lab tech began preventative medical treatment for (b)(6).Manufacturer testing was completed on july 08 for the retained samples as well as returned unused product from the customer.Testing occurred at (b)(4), an affiliate of thermofisher scientific, on the abbott architect (b)(6) assay and the procleix ultrio combo system.The results displayed no (b)(6) for (b)(6) from both systems.The customer sent the product for western blot testing with (b)(6) results for (b)(6) on all bands.The date for the testing is unk.No product malfunction has been asserted or identified.
|