• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL SYSTEMS AB PRESSUREWIRE AERIS; GUIDEWIRE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL SYSTEMS AB PRESSUREWIRE AERIS; GUIDEWIRE SENSOR Back to Search Results
Model Number C12058
Device Problem Device Handling Problem (3265)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/11/2014
Event Type  malfunction  
Event Description
The pressure-wire became stuck in the stent and the coil portion of the distal tip of the pressure-wire was found to be stretched.Another pressure-wire was used to successfully perform ffr measurement.According to the physician there was not any resistance or difficulty in delivering.
 
Manufacturer Narrative
No evidence of externally induced markings (like scratching), inherent weak spots (like inclusions) or flaws from the manufacturing process that could have contributed to the failure were found on the core wire.From the visual examination and analysis of sem images it is likely that the fracture was initiated on upper bend of the surface on typical grinding marks probably due to high rotational force and then propagate along the surface to the bottom of the fracture surface due to rotational force before it completely breaks by a small tensile force.A review of the device history record confirmed that the device was manufactured according to sjm specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSUREWIRE AERIS
Type of Device
GUIDEWIRE SENSOR
Manufacturer (Section D)
ST. JUDE MEDICAL SYSTEMS AB
palmbladsgatan 10
box-6350
uppsala SE-75 1 35
SW  SE-751 35
Manufacturer Contact
emil anderas
palmbladsgatan 10
box 6350
uppsala SE-75-1 35
SW   SE-751 35
8161000
MDR Report Key4005277
MDR Text Key20779255
Report Number8030904-2014-00012
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K113584
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/25/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/31/2016
Device Model NumberC12058
Device Lot Number141484
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/25/2014
Initial Date FDA Received07/24/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-