MAUDE Adverse Event Report: DEROYAL INDUSTRIES DEROYAL HOT/COLD PACK, 4.5IN X 9.5IN, 20/CS; PACK, HOT OR COLD, DISPOSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEROYAL INDUSTRIES DEROYAL HOT/COLD PACK, 4.5IN X 9.5IN, 20/CS; PACK, HOT OR COLD, DISPOSABLE Back to Search Results
Catalog Number 9380-20
Device Problem Product Quality Problem (1506)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/18/2014
Event Type  malfunction  
Event Description
The ice packs are not only leaking but the holes these are developing are 1/4" or bigger.These ice packs are spilling out onto the patient, getting the dressing soaked, and even getting near or on the patient's wound.
 
Manufacturer Narrative
Investigation findings: no sample was returned.During a review of the information listed in this qfi, i found that the product was being purchased from a distributor.With that being said, i reviewed the routing of the product - first the raw materials are pulled at the fab facility and shipped to (b)(4).At (b)(4) the packs are counted out and 20 are packaged/labeled then shipped (b)(6).Once the order is placed the finished good is pulled/billed and shipped to, in this case cardinal.Cardinal gets an order and ships the product to the facility that is filing the complaint.The facility is then placing the product on shelves in a freezer, before it is placed on the patient.These items are bright blue in color any leaks should be quite easily seen during handling.We are unsure if the product was damaged prior to having been used.The vendor was informed about the issues and requested samples, however, we were unable to provide any.They are unwilling to take responsibility for the damaged packs as there were no issues found during deroyal handling/review.We will be getting future orders from our old vendor as the facility has been reopened.Correction: replacement sent.Root cause analysis: possible enduser issues.Corrective action and/or systemic correction action taken: no further action is required as no vendor defect/error could be found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEROYAL HOT/COLD PACK, 4.5IN X 9.5IN, 20/CS
Type of Device
PACK, HOT OR COLD, DISPOSABLE
Manufacturer (Section D)
DEROYAL INDUSTRIES
164 giles hollow rd
rose hill VA 24281
Manufacturer Contact
200 debusk lane
powell, TN 37849
8653622333
MDR Report Key4009150
MDR Text Key4691982
Report Number1123071-2014-00016
Device Sequence Number1
Product Code IMD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 07/11/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number9380-20
Device Lot NumberMH73912
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date06/18/2014
Event Location Hospital
Initial Date Manufacturer Received 06/18/2014
Initial Date FDA Received07/14/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-