JOHNSON & JOHNSON MEDICAL (SUZHOU) LTD. 3006356043 PINN CAN BONE SCREW 6.5MMX35MM; HIP OTHER IMPLANT Back to Search Results
Catalog Number 121735500
Device Problem Break (1069)
Patient Problem No Information (3190)
Event Date 07/29/2014
Event Type  malfunction  
Manufacturer Narrative
The device associated with this report was not returned.A review of the device history records for the provided product and lot combination did not reveal any related manufacturing deviations or anomalies.Requests for additional investigational inputs were made in accordance with wi-7915 appendix a.No additional information was obtained.The investigation could not verify or identify any product contribution to the reported event with the information provided.Based on the inability to identify root cause, the need for corrective action was not indicated.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Event Description
Upon tightening the pinnacle screw into the pinnacle cup, the screw head broke.The head was removed from the patient and the screw shank was impeded in the bone but clear of the cup.A second screw was placed in a separate hole without incident and the polyethylene liner was placed in the shell without incident.Delay of 3 minutes.No ae to patient.
 
Manufacturer Narrative
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN CAN BONE SCREW 6.5MMX35MM
Type of Device
HIP OTHER IMPLANT
Manufacturer (Section D)
JOHNSON & JOHNSON MEDICAL (SUZHOU) LTD. 3006356043
no.299 changyang st
suzhou industrial park
suzhou, jiangsu 2151 2-6
CH  21512-6
Manufacturer (Section G)
JOHNSON & JOHNSON MEDICAL (SUZHOU) LTD. 3006356043
no.299 changyang st
suzhou industrial park
suzhou, jiangsu 2151 2-6
CH   21512-6
Manufacturer Contact
steve dowell
700 orthopaedic drive
warsaw, IN 46581
5743714918
MDR Report Key4011984
MDR Text Key12799408
Report Number1818910-2014-25701
Device Sequence Number1
Product Code NDJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK983014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/29/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number121735500
Device Lot NumberD00446342
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/30/2015
Initial Date FDA Received08/15/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/30/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-