• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMMUCOR, INC. CAPTURE-R SELECT; SOLID PHASE SYSTEM FOR THE IMMOBILIZATION OF HUMAN ERYTHROCYTES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMMUCOR, INC. CAPTURE-R SELECT; SOLID PHASE SYSTEM FOR THE IMMOBILIZATION OF HUMAN ERYTHROCYTES Back to Search Results
Catalog Number 0006446
Device Problem False Negative Result (1225)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/18/2014
Event Type  malfunction  
Event Description
On (b)(6) 2014 an immucor field personnel reported false negatives on an igg cross-match performed on galileo echo instrument with capture-r select lot sc280.
 
Manufacturer Narrative
Immucor technical support used remote electronic access method to assess the instrument electronic images.Sample in question reacted negative with multiple lots of capture reagents and different unit samples.Could not rule out sample related issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTURE-R SELECT
Type of Device
SOLID PHASE SYSTEM FOR THE IMMOBILIZATION OF HUMAN ERYTHROCYTES
Manufacturer (Section D)
IMMUCOR, INC.
3130 gateway drive
norcross GA 30091 562
Manufacturer (Section G)
IMMUCOR, INC.
3130 gateway drive
norcross GA 30091 562
Manufacturer Contact
howard yorek
3130 gateway drive
norcross, GA 30091-5625
7704412051
MDR Report Key4013410
MDR Text Key4708452
Report Number1034569-2014-00151
Device Sequence Number1
Product Code KSS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK040024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Medical Technologist
Type of Report Initial
Report Date 08/14/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Medical Technologist
Device Expiration Date05/05/2015
Device Catalogue Number0006446
Device Lot NumberSC280
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/18/2014
Initial Date FDA Received08/15/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/25/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age43 YR
-
-