Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/17/2014
Event Type  malfunction  
Event Description
Customer reported that instrument displayed ".? (question mark)" for electrolytes when patient sample run on it.There was no report of serious injury due to this event.
 
Manufacturer Narrative
As per urgent field safety notice 31512 rev a., this serial number cartridge was identified as one of that could exhibit this error.The customer should replace the measurement cartridge with a serial number after the cut off specified in the ufsn.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDPOINT 500
Type of Device
RP 500
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
steven andberg
2 edgewater drive
norwood, MA 02062
7812693655
MDR Report Key4017665
MDR Text Key20756924
Report Number1217157-2014-00120
Device Sequence Number1
Product Code KHP
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K113216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date 07/24/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Health Care Professional
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/24/2014
Initial Date FDA Received08/18/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-2079-2014
Patient Sequence Number1
-
-