MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE Back to Search Results
Model Number 3058
Device Problems Unable to Obtain Readings (1516); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/30/2014
Event Type  malfunction  
Event Description
It was reported that there was a possible issue with the implantable neurostimulator (ins).There was an impedance issue of the ¿???¿ error and all connections were blacked out with the lead being properly aligned and connected to the battery.It was noted that the action required as a result of the event was that a different product was used.The diagnostic testing performed was impedance testing.It was stated that the product issue was resolved and the cause of the issue was not determined.The health care provider (hcp) believed there was a malfunction with the ins and wanted a new ins.The new ins was opened and used for implantation and the ins that was never implanted would be returned.It was reported that the patient was alive with no injury and there were no patient symptoms or complications associated with the event.Additional information received reported that the patient was doing well and was receiving effective therapy.The device was being shipped back to the manufacturer today and a delay was experienced in waiting for the proper box to be received.
 
Manufacturer Narrative
(b)(4).
 
Manufacturer Narrative
Final analysis of neurostimulator model 3058 (lot # njy260874h ) showed no anomaly found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM II
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central avenue ne rcw215
minneapolis MN 55432
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key4024602
MDR Text Key18569374
Report Number3004209178-2014-15608
Device Sequence Number1
Product Code EZW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/30/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/28/2015
Device Model Number3058
Device Catalogue Number3058
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/26/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/22/2014
Initial Date FDA Received08/20/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received09/23/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/02/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age00066 YR
-
-