The referenced otv-s190 was returned to olympus medical systems corp (omsc) for eval.Omsc evaluated the otv-s190 and the endoscope which was used in conjunction with the otv-s190 and found that the abnormal image and/or noisy image were displayed, also the error appeared on the monitor.Furthermore, the locking lever of the otv-s190 was deformed and there were a lot of lint on the electrical contacts of the video connector.Olympus attributed this phenomenon to inappropriate handling of the device by user.The otv-s190 instruction manual states the notice for the handling of the device.Also, omsc checked the mfr history of the subject device, there was no irregularity found.There were no further details provided.If significant add'l info is received, this report will be supplemented.This report is being submitted as a medical device report in an abundance of caution.
|