During a transanal endoscopic procedure the co2 ran out.Once co2 was replaced the value on the device did not change and/or reset.All tubing was disconnected and then reconnected and an error 11 appeared on the device.Tubing was replaced with new, device turned off/on and error 11 still displayed on device.Delay in procedure may have placed patient at risk.Doctor proceeded on and opened up abdomen and performed a different procedure.Sometime while patient was on table, bowel was perforated.
|
An investigation was completed as the actual device was returned to the rwmic facility on (b)(4) 2014.Device history: sold on 01/12/2010, repaired 02/28/2011 and 06/2013.Unable to replicate issue.Device to be turned on and automatic function check performed prior to adding tubing to device.If tubing in device during automatic function check an error message occurs.Labeling was reviewed and found to be adequate, i.E.Intended use, indications and field of use, preparation and cautions.Error 11 is overpressure alarm, greater than 30 mm hg for more than 5 seconds.Richard wolf considers this matter closed.However, in the event we receive additional information, we will provide fda with follow-up information.
|