Model Number 20
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/02/2014
Event Type  malfunction  
Event Description
It was reported to physio-control that the customer's device during cardioversion on a patient automatically started the self-test again several times.The device kept rebooting and could therefore not deliver therapy.A back-up device was used to treat the patient.There was patient use associated with the reported event; however there was no adverse patient outcome reported.
 
Manufacturer Narrative
(b)(4).Physio-control evaluated the device and verified the reported failure.Physio-control continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Manufacturer Narrative
Physio-control observed that the device had logged multiple event codes which were indicative of a failure of the system pcb assembly and/or user interface pcb assembly.Physio-control advised replacement of the system pcb and user interface pcb assemblies.The customer declined repair of the device and requested to have the device returned without being repaired.A conclusive cause of the reported failure could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer Contact
jason march
4258674000
MDR Report Key4050308
MDR Text Key4864648
Report Number3015876-2014-01022
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K063119
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 08/04/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number3202488
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/07/2014
Is the Reporter a Health Professional? Yes
Device Age10 YR
Event Location Hospital
Initial Date Manufacturer Received 09/12/2014
Initial Date FDA Received09/01/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received09/17/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/15/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age72 YR
Patient Weight86
-
-