(b)(6).Complaint conclusion: the device and the echelon 10 microcatheter were returned for analysis.Proximal, but adjacent to the ball tip, the distal section of the coil has been kinked.The radius on the ball tip is distorted.The proximal end of the coil is buckled.Contrast-like material was found at the distal tip of the microcatheter, but the this was not significant.The microcatheter was flushed and a 0.014¿ guide wire was passed through without any resistance.Using multiple new cpl 100152-30 deltaplush, the test coils found a gap between the hub¿s funnel and the microcatheter¿s tube and became stuck.The damage at the distal tip of the test coils is similar to the returned coil¿s damaged distal section.The complaint event was duplicated using the returned microcatheter.Multiple test coils of the same catalog number became stuck just before the entrance into the microcatheter¿s tube.Therefore; the most likely root cause of the coil becoming stuck inside the microcatheter was due to a gap between the hub¿s funnel and the microcatheter¿s tube.This caused the distal tip of the coil to become stuck and damaged.Since there was no indication of a manufacturing issue related to the deltaplush product, no corrective action was taken.This mdr is being submitted as part of a retrospective review as the result of a recent fda audit and with accordance to the requirements of code of federal regulations - 21 cfr part 803, medical device reporting.
|