The pressure monitoring product that is related to this complaint was custom designed specifically for one hospital that is located outside of the united states.This product is not similar to any product that is used on patients inside of the united states.This hospital returned all unused product inventory to edwards lifesciences.An investigation ensued after notification of this event.A sr.Director of engineering and a sr.Manager of product safety from edwards visited the hospital located in the netherlands to receive firsthand knowledge of the event.The concerns expressed by the customer were: (1) the manifold becomes loose from the dpt side when manipulated, (2) the manifold becomes loose from the patient side when manipulated, and (3) the connection to the pressure tubing at the dpt becomes loose when manipulated.To examine these concerns, 20 models from the finished goods inventory were examined.Loose connections were detected in 1 out of 20 units at the patient side connector to the manifold.None of the other connections on these units were loose.The unit with the loose connection initially failed the leak testing, but after proper attachment of the connection, the unit passed the leak test.Additional investigation of the units found that 3 out of 20 manifolds had increased friction of the ¿rotating nut¿ on the patient side connection when the nut was pulled forward.This information was shared with the supplier of the manifold.All components were inspected and all components met specifications per their applicable drawings.In addition, 100 units in the next production lot were inspected for the measurements of the gap between the male luer connector on the pressure tubing and the rotating nut on the manifold.The results showed that there was a significant amount of variation in the tightening of that connection in this model.Based on the discussions and investigation results, it can be confirmed that the components met all specifications.
|
It was reported that the pressure monitoring set with manifold was used to monitor the cvp measurements and administer medication through the dpt system.The patient was administered noradrenaline and milrinone through the dpt system.The system was changed out after 72 hours.One hour later, the patient started having blood pressure fluctuations from 45-250mmhg systolic.The dosage of the medication was changed, but there was no improvement of the hemodynamic status.The clinician checked the dpt system and discovered the manifold was disconnected on the sensor side of the manifold.Estimated time of detachment was 1 minute.The patient was stabilized.No blood loss was reported.
|