It was reported per field service report: system - circuit breaker tripped during transport.Reason for service: check out.Findings/action taken: performed functional testing and was unable to duplicate customer's complaint.Unit passed a power supply voltage checkout and battery load test.The circuit breaker and power cord were replaced as a precaution.Unit passed functional checkout with electrical safety test.The pump is back in service.Additional information received on (b)(4) 2014 reported that the biomed stated to the field service representative the problem occurred after the transport was completed.The biomed didn't know if the patient was on the pump at the time of the incident.An update received on (b)(4) 2014 from the sales representative stated that there was no effect to the patient's condition.During transport the internal battery failure alarm occurred and requested to check circuit breaker.The pump was initially able to be run on ac power and they were able to transition the patient from the cardiac cath lab to the ambulance.The pump ran during transport to the other facility.On arrival to the other facility, the iabp would not turn on even when connected to ac power.Iabp taken oos (out of service), spare iabp placed in med 9.
|