Model Number N/A
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
The customer reported that an elevator broke during surgery.No adverse effects were reported as a result of this event.
 
Manufacturer Narrative
A review of the non-conformance database was conducted and there were no documented deviations for this lot.The package insert states "the tip of the instrument is extremely thin and delicate, care should be taken to avoid applying significant pressure to the tip.If excessive pressure is applied to the tip of the instrument, it will bend, rendering it useless.This elevator is not intended for use when removing a tooth that is not loose in the socket." upon evaluation, the tip of the instrument was found to be chipped and the fragmented piece was not returned.There were no indications of manufacturing defects.The most likely underlying cause of the complaint issue is excessive force applied by the user.There are no adverse effects reported as a result of this event, however, the device malfunction is subject to the two year rule.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEVATOR #301
Type of Device
ELEVATOR #301
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer (Section G)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer Contact
amanda sisk
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key4068664
MDR Text Key4850976
Report Number0001032347-2014-00306
Device Sequence Number1
Product Code EMJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 08/15/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number09-0257
Device Lot Number110112J12
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/21/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/15/2014
Initial Date FDA Received09/08/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-