|
Model Number H749236310030 |
Device Problem
Air Leak (1008)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/11/2014 |
Event Type
malfunction
|
Event Description
|
Reportable based on device analysis completed on (b)(4) 2014 same case as mdr id: 2134265-2014-05299.It was reported that a leak occurred.The stenosed target lesion was located in the moderately tortuous and severely calcified vessel.A 1.50mm rotalink¿ plus and 330cm rotawire¿ and wireclip¿ torquer were used to treat the target lesion.During procedure, a leaking sound was heard from the advancer.The procedure was completed with another of the same device.There were no patient complications reported and the patient¿s condition is good.However, device analysis revealed a tear in the sheath around the coil and that saline was being expelled through the sheath.
|
|
Manufacturer Narrative
|
Age at time of event: 18 years or older.Device evaluated by manufacturer: device was returned for analysis.A visual examination of the complaint unit was carried out.The advancer knob was returned in a backward position, it was loosened and advanced in order to inspect the handshake connection.The handshake connection was inspected and no damage was noted.A handshake connection test was attempted to examine the integrity of the connection and no issues were noted.The coil was observed to be kinked upon inspection.The complaint unit was wet tested and upon inspection it was noted there was a tear in the sheath around the coil and that saline was not reaching the end of the sheath but was being expelled through the sheath.As a result the device could not be successfully wet tested.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is considered handling damage as the event occurred without direct patient contact.(b)(4).
|
|
Search Alerts/Recalls
|
|
|