Model Number 20
Device Problem Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Date 06/30/2014
Event Type  malfunction  
Event Description
The customer reported that their device initially was showing its service indicator prior to being used on a patient.After an evaluation of the device by physio-control, it was observed that the device would lock up with a solid white screen upon attempting to power the device on.There was no patient use associated with the reported failure.
 
Manufacturer Narrative
(b)(4).Physio-control evaluated the device and verified the reported failure.Physio replaced the system controller and user interface pcb assemblies and observed proper device operation through functional and performance testing.The device was then returned to the customer for use.The removed pcb assemblies were further evaluated in the failure analysis center (where the lockup was initially observed), and it was confirmed that the cause of the reported failure was due to a failure of an integrated circuit chip, designator u61 from the system controller pcb assembly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer Contact
jason march
4258674000
MDR Report Key4082253
MDR Text Key4742207
Report Number3015876-2014-01074
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063119
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 08/15/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number3202488
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/18/2014
Is the Reporter a Health Professional? Yes
Device Age10 YR
Event Location Hospital
Initial Date Manufacturer Received 08/15/2014
Initial Date FDA Received09/11/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/28/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-