|
Model Number H749518080 |
Device Problem
Kinked (1339)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/25/2014 |
Event Type
malfunction
|
Event Description
|
Reportable based on device analysis completed on 21aug2014.It was reported that shaft kinked occurred.During the preparation to perform a percutaneous coronary intervention, an opticross imaging catheter was used to diagnose an unknown lesion.When setting up this device, the proximal shaft was kinked.The device was exchanged to another of same device to continue the procedure.No patient complications reported and the patient's condition is good.However, device analysis revealed an open hole at the sheath lap joint section of the device.
|
|
Manufacturer Narrative
|
(b)(4).Device evaluated by manufacturer: the complaint device was received for evaluation.Evaluation of the returned device revealed kinks in the sheath assembly from femoral marker to the distal end.The telescope assembly was not able to properly pull back, advance, or retract.Since the telescope cannot advance the transducer distal housing (tdh) to the most distal position, the distance from the distal end of the transducer housing to the tip of the catheter was not measured.An open hole was observed at the sheath lap joint section of the device.Fluid was leaking from the open hole at the sheath lap joint assembly when the catheter was flushed.The imaging window was still connected to the blue sheath tubing at the lap joint.No other issues or defects were observed during product analysis of the returned device.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|
|
Search Alerts/Recalls
|
|
|