• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER Back to Search Results
Catalog Number 1012455-15
Device Problems Detachment Of Device Component (1104); Difficult to Remove (1528); Folded (2630)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/21/2014
Event Type  malfunction  
Event Description
It was reported that using a brachial artery access approach during a procedure of the eccentric, 70% restenosed, non-tortuous, proximal left anterior descending (lad) artery predilatation with a 5.0 x 10 mm non-abbott balloon dilatation catheter (bdc) was completed at 6 atmosphere (atm) for 6 seconds and inflated again.The bdc was removed and the 4.0 x 18 mm xience prime stent was deployed at 14 atm for 30 seconds.It was noted that post-dilatation was completed using a 5.0 x 15 mm nc trek bdc at 12 atm and 16 atm but after deflation the bdc met resistance while attempting to be removed.The bdc was inflated and deflated twice and was removed with resistance.Although the indeflator was not on negative pressure, after removal from the anatomy the balloon was noted to be winged.Additionally, the bdc shaft was noted to have separated during removal.There was no reported adverse patient effect.There was no reported clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
(b)(4).The device was received.Investigation is not yet complete.A follow up report will be submitted with all relevant information.
 
Manufacturer Narrative
(b)(4).Evaluation summary: the device was returned for evaluation.The reported separation and winged balloon was confirmed.The reported difficult to remove the balloon catheter was not confirmed due to the condition of the returned device.Based on visual analysis of the returned device, there is no indication of a product deficiency.A review of the lot history record revealed no non-conformances that would have contributed to this complaint.A query of the complaint handling database revealed no other similar incidents reported from this lot.Based on the information reviewed, there is no indication of a product deficiency.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Type of Device
CORONARY DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer (Section G)
EL COYOL, COSTA RICA REG# 3009031392
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key4093375
MDR Text Key4714534
Report Number2024168-2014-05956
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K110134
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/22/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2016
Device Catalogue Number1012455-15
Device Lot Number40602G1
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/10/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/19/2014
Initial Date FDA Received09/16/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/07/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
STENT: XIENCE PRIME 4.0X18MM
Patient Age75 YR
-
-