Rn safety report:
situation: at 0100 care time air found to be leaking around ventilator tubing, no vent alarms, pt getting tidal volumes without increasing peak pressures.Background: ega 26 week on vent support.Action: rt called to bedside, vent circuit changed, pt provided ppv with neotee during change over.Response: tubing saved for rt.Rn: when repositioning infant steady stream of air felt flowing from vent tubing close to y connector.The vent was not alarming.Pt getting adequate tidal volumes with pressures of 15-18 which was consistent with earlier in shift.Leak registered at 15-25%.Vent circuit changed and leak consistent.No change to baby's breath sounds, work of breathing pressure demands to meet tidal volumes and o2 reads stable throughout.Respiratory therapist(rt): vent circuit tubing audibly leaking, no alarms, patient's tidal volume remained within normal limits, no increase in peak pressure noted.Circuit tubing changed out while baby was neopuffed.Hole in tubing found near inspiratory connection.Rt also stated it could have been caused by the vent being too close to the isolette.There was a note placed on the circuit tubing that was removed that said "hole discovered on exhalation side where the tape was placed.".
|