• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIEBEL FLARSHEIM HUT EXT DR FINAL ASSY-REVERSE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIEBEL FLARSHEIM HUT EXT DR FINAL ASSY-REVERSE Back to Search Results
Model Number DR FINAL A
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/11/2014
Event Type  malfunction  
Event Description
Customer reports via phone that during a urology stent placement, on a (b)(6) year old male patient the system fluoro failed.Physician completed the procedure using endoscopy without further incident.No reported injury.
 
Manufacturer Narrative
Field service engineer (fse) went on site, troubleshot the no fluoro issue to the generator console.Fse replaced the console, tested the unit for proper operation and returned the unit to full service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUT EXT DR FINAL ASSY-REVERSE
Manufacturer (Section D)
LIEBEL FLARSHEIM
2111 east galbraith rd.
cincinnati OH 45237
Manufacturer Contact
david benson,
2111 east galbraith rd.
cincinnati, OH 45237
5139485719
MDR Report Key4102310
MDR Text Key4720030
Report Number1518293-2014-00099
Device Sequence Number1
Product Code IXR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/11/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDR FINAL A
Device Catalogue Number404007
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/11/2014
Initial Date FDA Received08/27/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age45 YR
-
-