• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH BATT F/PD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH BATT F/PD Back to Search Results
Catalog Number 530.200
Device Problem Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
Customer reported the device gets very warm in the battery case.Probably internal closed circuit.Device was returned for service.This is report 2 of 3 for complaint #(b)(4).
 
Manufacturer Narrative
Additional narrative: synthes is submitting this report as a result of remediation activities related to synthes service and repair files legacy review/remediation protocol-complaint handling and mdr reporting.Device listed in this report is used for treatment, not diagnosis.Any additional information received regarding this event after filing this report shall be filed on a supplemental mdr.During pre-repair assessment performed by a technician, functional testing was performed and revealed the control unit was not functioning.This was attributed to normal wear.The device was repaired and returned to the customer on (b)(4) 2012.Placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATT F/PD
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
ch-4436
oberdorf
SZ 
Manufacturer (Section G)
SYNTHES GMBH
eimattstrasse 3
ch-4436
oberdorf
SZ  
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
8006207025
MDR Report Key4102826
MDR Text Key4840379
Report Number8030965-2014-01460
Device Sequence Number1
Product Code MOQ
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 08/22/2012
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.200
Device Lot NumberMC091216
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/22/2012
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/22/2012
Initial Date FDA Received09/19/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/14/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-