Event Description
The customer, a biomedical engineer, contacted physio-control to report that their device had a service indicator present.There was no patient use associated with the reported event.Upon evaluation of the customer's device, physio observed that the unit would constantly reboot itself upon power on and never complete the boot up cycle.
 
Manufacturer Narrative
Physio-control examined the customer's device and verified the reported issue.Physio then replaced the therapy pcb assembly, as well as made other unrelated repairs, and after observing proper device operation through functional and performance testing the unit was returned to the customer for use.Physio further evaluated the removed therapy pcb assembly and determined that the cause of the reported issue was an integrated circuit (ic) chip, (b)(4) was temperature sensitive (heat).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer Contact
jason march
4258674000
MDR Report Key4110244
MDR Text Key4836458
Report Number3015876-2014-01108
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063119
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 08/25/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number3202487
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/29/2014
Is the Reporter a Health Professional? Yes
Device Age11 YR
Event Location Hospital
Initial Date Manufacturer Received 08/25/2014
Initial Date FDA Received09/23/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/24/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-