PHYSIO-CONTROL, INC LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR; DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)
|
Back to Search Results |
|
Model Number 20 |
Device Problem
Device Inoperable (1663)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/21/2014 |
Event Type
malfunction
|
Event Description
|
The customer, a biomedical engineer, contacted physio-control to report that their device had a service indicator present.There was no patient use associated with the reported event.Upon evaluation of the customer's device, physio observed that the unit would constantly reboot itself upon power on and never complete the boot up cycle.
|
|
Manufacturer Narrative
|
Physio-control examined the customer's device and verified the reported issue.Physio then replaced the therapy pcb assembly, as well as made other unrelated repairs, and after observing proper device operation through functional and performance testing the unit was returned to the customer for use.Physio further evaluated the removed therapy pcb assembly and determined that the cause of the reported issue was an integrated circuit (ic) chip, (b)(4) was temperature sensitive (heat).
|
|