TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO PERFUSION SYSTEM 8000; 8K (PUMP, ROLLER TYPE)
|
Back to Search Results |
|
Model Number 16402 |
Device Problems
Sticking (1597); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/26/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the perfusionist (ccp) stated the occlusion knob was stuck in fully open position on the roller pump.As a result, an alternate device was employed.The surgical procedure was completed successfully.There were no delays, no blood loss, or no adverse consequences to the pt.
|
|
Manufacturer Narrative
|
The field service representative (fsr) used spanner wrench to loosen the occlusion knob.The fsr removed the occlusion knob, checked for burrs on knob and threaded shaft, none were found.The fsr removed the old grease, re-greased and re-installed occlusion knob.The fsr performed corrective/preventative maintenance on the roller pump.The occlusion knob moved freely and the unit operated to mfr specifications and was returned to clinical use.No parts will be returned to the mfr for eval.If additional info becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|