Model Number 16402
Device Problems Sticking (1597); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/26/2014
Event Type  malfunction  
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the perfusionist (ccp) stated the occlusion knob was stuck in fully open position on the roller pump.As a result, an alternate device was employed.The surgical procedure was completed successfully.There were no delays, no blood loss, or no adverse consequences to the pt.
 
Manufacturer Narrative
The field service representative (fsr) used spanner wrench to loosen the occlusion knob.The fsr removed the occlusion knob, checked for burrs on knob and threaded shaft, none were found.The fsr removed the old grease, re-greased and re-installed occlusion knob.The fsr performed corrective/preventative maintenance on the roller pump.The occlusion knob moved freely and the unit operated to mfr specifications and was returned to clinical use.No parts will be returned to the mfr for eval.If additional info becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO PERFUSION SYSTEM 8000
Type of Device
8K (PUMP, ROLLER TYPE)
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson rd.
ann arbor, MI 48103
7346634145
MDR Report Key4114249
MDR Text Key19589571
Report Number1828100-2014-00749
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K953901
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/26/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16402
Device Catalogue Number16402
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/26/2014
Initial Date FDA Received09/17/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-