• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR, LLC FUSION BIOLINE 8MM-60CM SUPP PERIPHERAL; VASCULAR GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR, LLC FUSION BIOLINE 8MM-60CM SUPP PERIPHERAL; VASCULAR GRAFT Back to Search Results
Model Number M00201503068B0
Device Problem Folded (2630)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/20/2014
Event Type  malfunction  
Event Description
The hospital report that during an implantation of a fusion bioline graft for a fempop.The surgeon commented that the graft folded while sewing anastomosis, this was the first time the surgeon implanted a fusion bioline.He did not think it would be an issue once he unclamped.No issue reported and he was able to close.
 
Manufacturer Narrative
Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.There was no conformance recorded in the lot history which would be considered related to the reported event.There are no other similar complaints reported against this batch.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION BIOLINE 8MM-60CM SUPP PERIPHERAL
Type of Device
VASCULAR GRAFT
Manufacturer (Section D)
MAQUET CARDIOVASCULAR, LLC
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR, LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer Contact
tina evancho
45 barbour pond drive
wayne, NJ 07470
9737097690
MDR Report Key4114766
MDR Text Key4699132
Report Number2242352-2014-01008
Device Sequence Number1
Product Code DSY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131778
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/21/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2016
Device Model NumberM00201503068B0
Device Lot Number25087424
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/21/2014
Initial Date FDA Received09/15/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-