MAUDE Adverse Event Report: SORIN GROUP DEUTSCHLAND S5 GAS BLENDER SYSTEM; GAS CONTROL UNIT, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP DEUTSCHLAND S5 GAS BLENDER SYSTEM; GAS CONTROL UNIT, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 25-28-67
Device Problem Other (for use when an appropriate device code cannot be identified) (2203)
Patient Problem No Patient involvement (2648)
Event Date 08/11/2014
Event Type  malfunction  
Event Description
Sorin group received a report that the s5 gas blender alarmed repeatedly during set up.There was no patient involvement.
 
Manufacturer Narrative
Sorin group (b)(4) manufacturers the s5 gas blender system.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Manufacturer Narrative
Sorin group (b)(4) manufactures the s5 gas blender system.The incident occured in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group received a report that the s5 gas blender alarmed repeatedly during setup.Investigation of the returned gas blender confirmed the reported issue.A defective mass flow controller for co2 caused the problem.The part was replaced.Testing, calibration and safety inspections were performed with no deviations.A review of the device history record did not find any concessions, deviations or non-conformities related to the reported failure.The issue will be monitored for trends and if identified, corrections will be recommended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 GAS BLENDER SYSTEM
Type of Device
GAS CONTROL UNIT, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr 25
munich D 809 39
GM  D 80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr 25
munich D 809 39
GM   D 80939
Manufacturer Contact
cheri voorhees, mgr
14401 west 65th way
arvada, CO 80004
3034676527
MDR Report Key4115736
MDR Text Key4703926
Report Number1718850-2014-00321
Device Sequence Number1
Product Code DTX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K101046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,health professional
Reporter Occupation Unknown
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/11/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number25-28-67
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/11/2014
Initial Date FDA Received09/09/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/20/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-