DEPUY SYNTHES POWER TOOLS 11.25CM ATTACHMENT, BM; BUR, SURGICAL, GENERAL & PLASTIC SURGERY - ATTACHMENT
|
Back to Search Results |
|
Catalog Number B-BLUE-S_LL |
Device Problem
Overheating of Device (1437)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/05/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that during pre-surgery, it was observed that the attachment had heat.There were no delays to the planned surgical procedure as an identical spare device was available for use.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
Additional narrative: the actual device was returned for evaluation.Reliability engineering evaluated the device and observed that a cutter could not be inserted into the device due to worn bearings.It was determined that worn and loose bearings usually create a situation where the cutter is not able to be inserted due to bearings that are no longer in axial alignment.It was determined this would have caused heat if a cutter was able to be inserted and the device was run.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to normal wear and servicing over time.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
|