• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IRIS INTERNATIONAL IQ200; AUTOMATED URINE MICROSCOPY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IRIS INTERNATIONAL IQ200; AUTOMATED URINE MICROSCOPY ANALYZER Back to Search Results
Catalog Number 700-7177-001
Device Problems Charred (1086); Device Emits Odor (1425); Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/01/2014
Event Type  malfunction  
Event Description
Customer reported burning smell from the instrument.
 
Manufacturer Narrative
A burning smell was reported from the instrument.Field service engineer (fse) found ssmb pcb slot no 8, 9 and the backplane pcb burned out.Fse replaced the necessary pcbs, calibrated and tested controls for iq and ichem units and all passed.Customer ran 30 abnormal samples without any issues.No injuries, any visible flames, or smoke emission were reported and the fire department was not called.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IQ200
Type of Device
AUTOMATED URINE MICROSCOPY ANALYZER
Manufacturer (Section D)
IRIS INTERNATIONAL
9172 eton ave.
chatsworth CA 91311
Manufacturer Contact
sudha gupta
9172 eton ave.
chatsworth, CA 91311
8185277272
MDR Report Key4137887
MDR Text Key4934090
Report Number2023446-2014-00154
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K022774
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/01/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number700-7177-001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/01/2014
Initial Date FDA Received09/29/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-