(b)(4).Upon receipt in our post market quality assurance laboratory, the device was interrogated by boston scientific's quality assurance laboratory and was found with a battery status indicator of end-of-life (eol) at 2.57 volts.Visual inspection noted that the device header was loose but had not separated from the device case.The device did not have good adhesion to the header.Additionally, tool marks were identified on the casing.An x-ray of the device verified that the feed-thru wires were stretched but not separated.The device passed manual therapy verification testing on the bench.The device was put through and passed manual diagnostic testing that verified the performance of pacing, sensing and shocking function of the device.The recorded pacing and shock impedance measurements were within normal limits.Manufacturing enhancements have been implemented to make the header bond more robust.
|