• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GE OEC MEDICAL SYSTEMS (SLC) 8800; FLUOROSCOPIC X-RAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GE OEC MEDICAL SYSTEMS (SLC) 8800; FLUOROSCOPIC X-RAY Back to Search Results
Model Number 8800
Device Problems Device Slipped (1584); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/28/2014
Event Type  malfunction  
Event Description
It was reported that the sys was having trouble starting up and staying up.This resulted in an intermittent, recoverable loss of imaging functionality.There is no report of injury or death associated with this event.
 
Manufacturer Narrative
A ge service representative performed an on site investigation.The reported issue could not be duplicated.The connectors were reseated during the service call.The system was tested and found to be working as intended and put back into service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
8800
Type of Device
FLUOROSCOPIC X-RAY
Manufacturer (Section D)
GE OEC MEDICAL SYSTEMS (SLC)
384 wright brothers dr.
salt lake city UT 84116
Manufacturer (Section G)
GE OEC MEDICAL SYSTEMS (SLC)
384 wright brothers dr.
salt lake city UT 84116
Manufacturer Contact
384 wright brothers dr.
salt lake city, UT 84116
8015364952
MDR Report Key4142931
MDR Text Key4931099
Report Number1720753-2014-07880
Device Sequence Number1
Product Code JAA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 09/12/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8800
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/28/2014
Initial Date FDA Received09/12/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-