Catalog Number 4115000000 |
Device Problems
Delayed Charge Time (2586); Battery Problem (2885)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/08/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that the batteries were not charging at user facility and there was a delay of one hour delay to the surgery.The procedure was completed successfully, no medical intervention and no adverse consequences were reported with this event.
|
|
Event Description
|
It was reported that the batteries were not charging at user facility and there was a delay of one hour delay to the surgery.The procedure was completed successfully, no medical intervention and no adverse consequences were reported with this event.
|
|
Manufacturer Narrative
|
The device has been received, failure analysis is in progress; additional information will be submitted once the quality investigation is completed.
|
|
Manufacturer Narrative
|
The reported event, batteries do not charge anymore, was not duplicated.The technician confirmed that the charger indicated ¿replace¿ when the batteries were place on the charger.All four batteries were able to be conditioned to normal capacity; however upon disassembly, all of the batteries had internal vented electrolytes.One battery also had internal corrosion.Devices are placed in parts retention.
|
|