• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. ANATOEMIC, 20 MM; PROSTHESIS, TOE, HEMI-, PHALANGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. ANATOEMIC, 20 MM; PROSTHESIS, TOE, HEMI-, PHALANGEAL Back to Search Results
Catalog Number AR-9500-200
Device Problem Insufficient Information (3190)
Patient Problems Erythema (1840); Pain (1994); Rash (2033); Swelling (2091); Reaction (2414)
Event Date 09/11/2014
Event Type  Injury  
Event Description
It was reported that a rash was first noted on the patient's foot near the implants on both toes and under the foot.The rash then traveled to both her hands.Follow-up information: the patient is having a reaction after an anatomic procedure of the 2nd and 3rd toes of her left foot.A rash was first noted on the patient's foot near the implants on both toes and under the foot.The blistering rash then traveled to both of her hands and wrist.The areas of the rash are red, painful and swollen.The patient had a basic allergy test done and was negative.She will now have a heavy metal patch test performed.
 
Manufacturer Narrative
Patient demographics (age at time of event, date of birth, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.The device was requested for evaluation but part remains in the patient and cannot be returned therefore the complainant's event could not be verified.The cause of the event could not be determined from the information available and without device evaluation.Device history record review revealed nothing relevant to this event.Based on the information provided, the most likely cause(s) of this event is an adverse reaction of the patient to the material(s) implanted.Product directions for use warns of adverse effects like foreign body and allergic-like reactions as well as infections both deep and superficial to the implant materials.Patient sensitivity to device materials must be considered prior to implantation.This is the first complaint of this type for this part/lot combination.If additional relevant information is received, a follow-up report will be submitted.Part remains in patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANATOEMIC, 20 MM
Type of Device
PROSTHESIS, TOE, HEMI-, PHALANGEAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 194
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 194
Manufacturer Contact
vik bajnath, sr. mdr analyst.
1370 creekside boulevard
naples, FL 34108-1945
8009337001
MDR Report Key4148017
MDR Text Key4916677
Report Number1220246-2014-00183
Device Sequence Number1
Product Code KWD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 09/11/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/01/2017
Device Catalogue NumberAR-9500-200
Device Lot Number1248011
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/11/2014
Initial Date FDA Received10/07/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-