Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.A visual inspection of the device header and case noted body fluid contamination in the lead barrels and a hole in the right atrial seal plug.Analysis noted that no shocks were delivered after the painless shock impedance measured less than 20 ohms.Analysis review of the device¿s memory found the device did shock into an open lead condition.No noise was noted during testing.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.
|