Customer reported via email: during bedside rcps initial patient and ventilator assessment, rcp noticed abnormally low exhaled tidal volumes. after quick trouble-shooting, rcp discovered a section of the expiratory limb on the ventilator circuit to be melted with small holes where the air was leaking out. fisher paykel heater was set at 37 degrees. no audible or visual alarms were activated on the ventilator nor the heater. the circuits were not touching, and no cloth material was surrounding the circuits.
|
(b)(4).Unfortunately, no sample was received for evaluation.A review of the documentation of internal manufacturing device history records could not be reviewed due to no lot number being reported.During the manufacturing process review, it was observed that product is 100% tested by resistance function.No assembly errors were observed during current production.In addition, no issues were found with the materials or design; the materials used for this catalog are validated.Based on the investigation, the most probable cause for the issue reported could not be determined without a sample.As a preventive action, carefusion recommends following the product label for recommended uses.The product label states the following applications:
humidifier wire adapter minimum flow rate
mr850 900mr805 3.0lpm
mr700, mr720, 900mr900 1.5 lpm
mr730, hc500
cautions: do not use this circuit where gas temperature at the outlet of the humidifier exceeds 68 degrees celsius.Do not use the heater circuit without gas flow.Do not place material on or around the heated wire tubing.Carefusion will continue to monitor and trend this issue.
|