• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - SPENCER RX CYTOLOGY BRUSH; ENDOSCOPIC CYTOLOGY BRUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - SPENCER RX CYTOLOGY BRUSH; ENDOSCOPIC CYTOLOGY BRUSH Back to Search Results
Model Number M00545000
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Event Description
Note: this event, as reported, did not reflect an mdr reportable scenario; however, evaluation of the returned device revealed an mdr-reportable malfunction.This manufacturer report is being submitted based on the evaluation results.It was reported to boston scientific corporation that an rx cytology brush was used in a brush cytology procedure.According to the complainant, during preparation, the brush failed to extend.It was suspected that the sheath was longer than usual.The procedure was completed with another rx cytology brush.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be good.
 
Manufacturer Narrative
Investigation result of wire broken.Analysis of the returned rx cytology brush did not confirm the reported complaint.The distal part of the catheter and the brush were cut off and not returned, therefore, it is not possible to determine if the working length was too long and if the brush was difficult to extend.Additionally, the drive wire was found to be bent and broken.Therefore the most probable root cause is operational context since residue was found on the entire device indicating that it was used during a procedure.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.The complaint was not confirmed.The device did not show any evidence of either the alleged issue or any defect which could have contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX CYTOLOGY BRUSH
Type of Device
ENDOSCOPIC CYTOLOGY BRUSH
Manufacturer (Section D)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
Manufacturer (Section G)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key4152994
MDR Text Key4932343
Report Number3005099803-2014-03267
Device Sequence Number1
Product Code FDX
Combination Product (y/n)N
PMA/PMN Number
K930348
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 09/14/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberM00545000
Device Catalogue Number4500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/05/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/14/2014
Initial Date FDA Received10/08/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-