On (b)(6) 2014, the patient underwent an endovascular procedure using a gore® excluder® aaa endoprosthesis featuring c3® delivery system to repair an abdominal aortic aneurysm.The patient's left external iliac artery was highly calcified, with the diameter of 3-6mm.It was reported that advancement of a gore® dryseal sheath with hydrophilic coating (dsl1228j/12778970) for implanting a contralateral leg component (pxc141400j/12616829) at the left side was not possible due to the calcification, and that a ballooning to the left external iliac artery did not help the advancement.The physician elected to advance the device outside the sheath, however the device was stuck at the aortic bifurcation and did not advance any further.The physician elected not to implant the device and during retracting the device it was stuck at the external iliac artery and then only the delivery catheter was eventually retracted.It was confirmed that the device was unintentionally deployed at the external iliac artery.The physician decided to leave the device in the artery.The patient tolerated the procedure.
|