Patient Problem Osteolysis (2377)
Event Type  Injury  
Event Description
It was reported that the patient underwent a dlif procedure using rhbmp-2/acs with an interbody device.Bmp was placed in the interbody device in the disc space.The surgeon reports that there was no endplate violation during the procedure, but the endplate may have been compromised due to osteoporosis.Approximately one year post-op, the ct and mri films showed some vertebral body osteolytic activity.
 
Manufacturer Narrative
(b)(4).Review of radiographic imaging found as follows: this patient underwent dlif at l4.Unilateral pedicle screws were used.Ap and lateral x-rays, ct axial and sagittal as well as mri are available for review.Small amount of edematous areas are noted around the screw tips and at the endplates above and below the clydesdale spacer.Small amount of osteolysis is noted in the anterior aspects of the endplates that are small enough not to be structural.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSE BONE GRAFT
Type of Device
FILLER, RECOMBINANT HUMAN BONE MORPHOGENETIC PROTEIN, COLLAGEN SCAFFOLD WITH MET
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK
1800 pyramid place
memphis TN 38132
Manufacturer Contact
huzefa mamoola
1800 pyramid place
memphis, TN 38132
9013963133
MDR Report Key4154982
MDR Text Key21494863
Report Number1030489-2014-03955
Device Sequence Number1
Product Code NEK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P000058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 09/09/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/09/2014
Initial Date FDA Received10/09/2014
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
INTERBODY SPACER
Patient Outcome(s) Other;
-
-