BOSTON SCIENTIFIC - GALWAY WALLSTENT? ENTERAL ENDOPROSTHESIS; STENT, COLONIC, METALIC, EXPANDABLE
|
Back to Search Results |
|
Model Number M00565590 |
Device Problems
Material Perforation (2205); Activation, Positioning or Separation Problem (2906)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/05/2014 |
Event Type
malfunction
|
Event Description
|
It was reported to boston scientific corporation that a wallstent enteral stent was used in the second part of the duodenum during a stent placement procedure performed on (b)(6) 2014.According to the complainant, the stent was used to treat a 5cm malignant stricture.Reportedly, the patient¿s anatomy was not tortuous.During the procedure, the physician attempted to deploy the stent when a strong resistance was felt and the outer sheath could not be retracted.The stent was removed from the patient and it was noted that 2 wires of the stent perforated the distal sheath.The physician cut the distal 2 cm tip of the catheter and reinserted the stent to patient but the stent was not able to be fully deployed.The stent was removed from the patient partially deployed 1-2cm.The procedure was completed with another wallstent enteral stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|
Manufacturer Narrative
|
A visual examination of the returned device found that the stent was received fully mounted onto the delivery system.It was noted that the tip section of the catheter had been removed prior to receipt and the t-bar connector was detached from the outer sheath.The stent was able to be deployed during analysis by manually retracting the outer sheath and a visual examination confirmed that the distal end of the stent was damaged.It was noted that some of the wires were uncrossed and raised vertically.This damage may have occurred during the stent wire perforation through the outer sheath and when the distal end of the device was cut.The dark blue outer sheath had fully detached from the valve body.This type of damage is consistent with excessive force being applied to the delivery system.Glue was attached to the t-bar connector indicating that it had been bonded to the outer sheath as required during manufacture.A visual and tactile examination confirmed that the dark blue outer sheath was kinked 16cm distal to the proximal end of the outer sheath.The outer sheath was also stretched and accordioned at several positions along its length.A visual and microscopic examination of the inner catheter identified several kinks along its length.There were no issues noted with the profile of the stent holder.No other issues were identified during the product analysis.The noted damages were likely due to procedural or anatomical factors encountered during the procedure such as tortuous anatomy or maneuvering of the device.Therefore, the most probable root cause for this complaint is operational context.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications.There was evidence that the device was used in a manner inconsistent with the labelled indications.The physician cut the distal tip off the delivery system and reinserted the device after discovering that the stent wires had perforated the outer sheath.The following precaution is stated in the dfu: ¿if sterility or performance of the device is suspected to be compromised, it should not be used¿.
|
|
Event Description
|
It was reported to boston scientific corporation that a wallstent enteral stent was used in the second part of the duodenum during a stent placement procedure performed on (b)(6) 2014.According to the complainant, the stent was used to treat a 5cm malignant stricture.Reportedly, the patient¿s anatomy was not tortuous.During the procedure, the physician attempted to deploy the stent when a strong resistance was felt and the outer sheath could not be retracted.The stent was removed from the patient and it was noted that 2 wires of the stent perforated the distal sheath.The physician cut the distal 2 cm tip of the catheter and reinserted the stent to patient but the stent was not able to be fully deployed.The stent was removed from the patient partially deployed 1-2cm.The procedure was completed with another wallstent enteral stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|
Manufacturer Narrative
|
(b)(4) sheath wire perforation.(b)(4) stent partially deployed.The device has been received for analysis; however the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
|
Search Alerts/Recalls
|
|
|