Reportable based on device analysis completed on 06oct2014.It was reported that a catheter kinked and outer telescope tubing detachment occurred.During a percutaneous coronary intervention (pci), an opticross¿ imaging catheter was selected to visualize an unspecified vessel.When the device was unpacked, it was noted that the guidewire exit port was kinked.Furthermore, the connection of the anchor seal housing was found detached without any physician contact on the device component.Moreover, it was also noted that the outer (female) telescope tubing between the anchor seal and the distal strain relief of the device was detached.The procedure was completed with another of the same device.No patient complications were reported and the patient¿s condition is good.However, device analysis revealed that the lap joint section of the device was damaged and a hole was observed.
|
(b)(4).The complaint device was returned for evaluation.Device analysis revealed that the lap joint section was damaged when the device was received.During handling of the device during the decontamination process, the imaging window became detached from the blue sheath assembly.No damage was observed at the guidewire exit port or at the distal tip assembly.No kinks were observed along the length of the catheter.The telescope female tubing was found detached from the anchor seal assembly.The telescope cannot advance the transducer distal housing (tdh) to the most distal position due to the detachments.Full image characterization cannot be performed due to the telescope and imaging window detachment.The detachment exposes the imaging core which might onset a windup when a functional test is performed.No other issues or defects were observed during product analysis of the returned device.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause was unable to be determined.(b)(4).
|