• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO CBCII(NO DRN)PKG/6 W/CAPACITOR; APPARATUS, AUTOTRANSFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO CBCII(NO DRN)PKG/6 W/CAPACITOR; APPARATUS, AUTOTRANSFUSION Back to Search Results
Catalog Number 0225028000E
Device Problem Coagulation in Device or Device Ingredient (1096)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/31/2014
Event Type  Injury  
Event Description
It was reported that at the user facility, clots formed, preventing transfusion of collected autologus blood from the device to a bag.The patient could not receive this blood and was administered two doses of blood from another source.
 
Manufacturer Narrative
The device was not returned for analysis.The device was not returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CBCII(NO DRN)PKG/6 W/CAPACITOR
Type of Device
APPARATUS, AUTOTRANSFUSION
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-PUERTO RICO
las palmas industrial park
highway #3, km 130.2
arroyo 0061 5
Manufacturer Contact
casey metzger
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key4202109
MDR Text Key5037122
Report Number0001811755-2014-03770
Device Sequence Number1
Product Code CAC
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
K952224
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Nurse
Type of Report Initial
Report Date 10/03/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0225028000E
Device Lot Number14173012
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/03/2014
Initial Date FDA Received10/27/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/22/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-