A review of lot c124 was performed and there were four non-conformances associated with this lot.None were related to the nature of this complaint.This lot met release requirements.A review of lot c124 shows no trends.A review of complaint categories shows no trend for "pressure dome membrane leak." capa (b)(4) and capa (b)(4) have already been initiated for further investigation of pressure dome membrane leaks.The returned kit was received for analysis.Examination of the system pressure dome showed evidence that it was not properly installed.When the pressure dome was cleaned and its diaphragm was fully seated, the pressure dome fit onto and stayed in place on a pressure sensor and a pressure test did not find any leaks.The reassembled system pressure dome performed as designed.Complaints are monitored through tracking and trending.If a trend is detected, further action will be taken.(b)(4).
|