• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE® LI-ION BATTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE® LI-ION BATTERY Back to Search Results
Model Number 8700-0752-01
Device Problems Break (1069); Positioning Problem (3009)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/06/2014
Event Type  malfunction  
Event Description
Complainant alleged that an autopulse® li-ion battery (s/n (b)(4)) would not stay seated in the autopulse® platform.However, when placed into the charger, the battery was able to be seated.It appears that one of the tabs in front of the battery was damaged.There were no issues when another battery was placed into the autopulse® platform.No adverse patient sequelae was reported.No further information was provided.
 
Manufacturer Narrative
Zoll has not yet received the autopulse battery in complaint for investigation.A supplemental report will be filed if and when the product is returned and investigation has been completed.
 
Manufacturer Narrative
The autopulse battery (s/n (b)(4)) was returned to the manufacturer and evaluated.The battery would not stay seated in a test autopulse platform.Visual inspection identified that one of the battery's right connector pins was damaged leading to the reported issue.Based on condition of the pin, the damage appears to have occurred during use and handling of the battery in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® LI-ION BATTERY
Type of Device
LI-ION BATTERY
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
joy patel
2000 ringwood ave,
san jose, CA 95131
4084192957
MDR Report Key4211929
MDR Text Key4986009
Report Number3010617000-2014-00572
Device Sequence Number1
Product Code DRM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 10/06/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8700-0752-01
Device Catalogue Number8700-0752-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/17/2014
Initial Date FDA Received10/29/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/22/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/11/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-