SORIN GROUP ITALIA S.R.L. - CRM FACILITY REPLY; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number REPLY DR
Device Problems Unable to Obtain Readings (1516); Failure to Transmit Record (1521)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/17/2013
Event Type  Injury  
Event Description
During a follow-up in (b)(6) 2013 (exact date not confirmed) it was not possible to open the episodes recorded in device memory despite using different programmers.
 
Event Description
During a follow-up in (b)(6) 2013 (exact date not confirmed), it was not possible to open episodes recorded in device memory despite using different programmers.
 
Manufacturer Narrative
It should be noted that this event was already submitted in 2013 (mdr#1000165971-2013-00541).Investigation concluded: normal device behavior could be restored by forcing the device to switch to standby mode before performing a re-initialization.It was reported that a new follow-up will be performed on (b)(6) 2014 and the recommendations will be applied.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REPLY
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
SORIN GROUP ITALIA S.R.L. - CRM FACILITY
parc d'affaires noveos 4 avenue réaumur
.
clamart 9214 0
FR  92140
Manufacturer (Section G)
SORIN GROUP ITALIA SRL VIA CRESCENTINO 13040 SALUGGIA ITALY
parc d'affaires noveos 4 avenue réaumur
.
clamart 9214 0
FR   92140
Manufacturer Contact
elodie vincent
parc d'affaires noveos 4 avenue réaumur
.
clamart 92140
FR   92140
0146013665
MDR Report Key4212417
MDR Text Key5238516
Report Number1000165971-2014-00602
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
PP950029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/29/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/25/2012
Device Model NumberREPLY DR
Device Catalogue NumberREPLY DR
Device Lot Number2516
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date10/29/2013
Event Location Hospital
Initial Date Manufacturer Received 12/19/2014
Initial Date FDA Received10/30/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/05/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/25/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-