Model Number REPLY DR |
Device Problems
Unable to Obtain Readings (1516); Failure to Transmit Record (1521)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/17/2013 |
Event Type
Injury
|
Event Description
|
During a follow-up in (b)(6) 2013 (exact date not confirmed) it was not possible to open the episodes recorded in device memory despite using different programmers.
|
|
Event Description
|
During a follow-up in (b)(6) 2013 (exact date not confirmed), it was not possible to open episodes recorded in device memory despite using different programmers.
|
|
Manufacturer Narrative
|
It should be noted that this event was already submitted in 2013 (mdr#1000165971-2013-00541).Investigation concluded: normal device behavior could be restored by forcing the device to switch to standby mode before performing a re-initialization.It was reported that a new follow-up will be performed on (b)(6) 2014 and the recommendations will be applied.
|
|