During placement of two 3m espe mdi one piece implant o-ball abutment 2.9 x 13mm for denture stabilization, the implants fractured and required surgical intervention to remove.The implants were being placed into tooth positions 21 and 23 in a patient with d2/d3 bone when they fractured at the o-ball region; at the time of fracture, a winged thumb wrench was being used for placement.The patient was referred to an oral surgeon who removed bone between the implants to remove them; bone grafting was performed.
|
It was noted that the dental professional was not using an adjustable torque wrench at the time the fracture occurred.In the 3m espe instructions for use, it is recommended to use an adjustable torque wrench, so that forces on the implant do not exceed 45 n- cm.It is possible that the forces applied using the winged thumb wrench (as was used in this case) exceeded the recommended 45 n-cm.
|