• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. ACL TIGHTROPE RT; PIN, FIXATION, SMOOTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. ACL TIGHTROPE RT; PIN, FIXATION, SMOOTH Back to Search Results
Catalog Number AR-1588RT
Device Problems Break (1069); Detachment Of Device Component (1104); Loose or Intermittent Connection (1371); Sticking (1597); Noise, Audible (3273)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/09/2014
Event Type  Injury  
Event Description
It was reported that an acl tightrope rt was seated and graft brought into the knee 5mm short of socket length.Tibia was fixed with screw and graft advanced again to further tension.When knee was cycled, there was a loud 'pop' and the graft was loose.The loop had broken off the button; it was removed and another implant was used to complete the case.There were no unplanned incisions but the original button remained in the patient.Follow-up: rep stated that the surgeon pulled the construct out after hearing the pop and the graft was loose but the button was stuck between the tissues and bone.A second device was inserted next to the 1st.One using the same insertion site.Pt.Is doing fine to date.
 
Manufacturer Narrative
Patient demographics (age at time of event, date of birth, gender, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.The device was requested for evaluation but was not returned, therefore the complainant's event could not be verified.The cause of the event could not be determined from the information available and without device evaluation.Device history record review revealed nothing relevant to this event.Based on the information provided, this type of event is most likely caused by applying excessive force on the shortening strands, nicking the suture with another instrument, fraying from sharp edge of the bone tunnel.The potential cause(s) of this event will be communicated to the event reporter.If additional relevant information is received, a follow-up report will be submitted.Facility will not return the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACL TIGHTROPE RT
Type of Device
PIN, FIXATION, SMOOTH
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 194
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 194
Manufacturer Contact
vik bajnath, sr. mdr analyst.
1370 creekside boulevard
naples, FL 34108-1945
8009337001
MDR Report Key4222526
MDR Text Key4995750
Report Number1220246-2014-00215
Device Sequence Number1
Product Code HTY
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K112990
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative,Distributor
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 10/16/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/01/2019
Device Catalogue NumberAR-1588RT
Device Lot Number1189787
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/16/2014
Initial Date FDA Received11/04/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-