• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOSPIRA COSTA RICA LTD. PCA PLS II; 80 MEA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOSPIRA COSTA RICA LTD. PCA PLS II; 80 MEA Back to Search Results
Catalog Number 01950
Device Problems Occlusion Within Device (1423); Failure to Sense (1559)
Patient Problem No Patient Involvement (2645)
Event Date 10/14/2014
Event Type  malfunction  
Event Description
The customer contact reported that during preventive maintenance testing at use facility, the device did not pass the occlusion test.There were no reports of any adverse pt events or delays in critical therapies while the device was in clinical use.No additional info was provided.
 
Manufacturer Narrative
The device was received.Investigation is not complete.This report represents all info known by the reporter upon query by hospira personnel.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PCA PLS II
Type of Device
80 MEA
Manufacturer (Section D)
HOSPIRA COSTA RICA LTD.
zona franca global
la aurora heredia
CS 
Manufacturer (Section G)
HOSPIRA COSTA RICA LTD.
zona franca global
la aurora heredia
CS  
Manufacturer Contact
juergen schmider, md vp gps
275 n field drive, bldg no h2-1east, dept no 097u
lake forest, IL 60045
2242125740
MDR Report Key4239257
MDR Text Key4988110
Report Number9615050-2014-05861
Device Sequence Number1
Product Code MEA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K912928
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 10/14/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number01950
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer10/17/2014
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/14/2014
Initial Date FDA Received10/31/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/01/1992
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-