The customer, a biomedical engineer, contacted physio-control to report that their device would only intermittently detect that a quik combo therapy cable was attached during testing.If the quik combo therapy cable is not detected, then a patient load (test or otherwise) may not be detected and defibrillation may not be possible, or delayed, if it were necessary.There was no patient use associated with the reported event.
|
(b)(4): physio-control provided the customer, a biomedical engineer, with technical assistance.It was later confirmed by the biomedical engineer that the therapy connector assembly was replaced which resolved the reported issue.After observing proper device operation through functional and performance testing the unit was placed back into service for use.The device has not been returned to physio-control for evaluation.
|